Revolutionize your performance by transforming your conventional, batch dynamic light scattering system into the stable, automated DynaPro™ Plate Reader™. The
DynaPro Plate Reader outperforms other DLS technologies by virtually every conceivable measure and provides unparalleled levels of ease-of-use, productivity, reproducibility, and flexibility for the most demanding applications.
Wyatt Technology is the world leader in providing the most advanced macromolecular and nanoparticle characterization tools. These instruments and software may be used to determine the absolute molecular weight and or size of macromolecules and nanoparticles and include: multi-angle light scattering, dynamic light scattering, high through-put dynamic light scattering, field flow fractionation, refractive index and viscometry detection.
Supported Well Plate Formats 96 or 384 or 1536 *
Size Range (radius) 0.5 to 1000 nm
Minimum Concentration @14 kDa (mg/mL) 0.125 †
Laser Wavelength (nm) 830 nm
Laser Power (mW) 7.5-75 mW (Programmable)
Attenuation Range 1 to 106
Temperature Control (C) 4º to 70ºC ††
Minimum Sample Volume 1.5 μL (using 1536 well plate)†††
Correlator (100 nsec sampling time 512 channels in a multi-tau layout)
Data Acquisition Time(s) 1 to 3600
Read Time per Well (average range) 5-20 seconds
Optical Fiber Receiver Proprietary Multi-Mode
Digital Communication One Standard Network port (TCP/IP)
Dimensions 59 cm (L) x 36 cm (W) x 21 cm (H);
23” x 14” x 8”
Total Weight 26 kg (instrument weight 18.5 kg)
Electrical 200 W, 90-250V @ 50-60 Hz,
universal power input
Specifications subject to change without notice.
* Supports most industry standard well plates.
† Minimum concentration specified with 50 μL of Lysozyme (14 kDa) sample in a 384 well plate, Wyatt P/N P8803-384,
with 100 seconds acquisition. Minimum concentration is inversely proportional to sample molar mass (i.e., a 28 kDa
sample may be measured at half the minimum concentration specified above.)
†† Absolute accuracy of +/– 1ºC from 4º-50ºC, absolute accuracy of +/– 1.5ºC from 50–70ºC. Dry inert gas required for
operation below 20ºC, at an operating pressure of 50–80 psi (0.34 - 0.55 MPa). At 60 psi the system will use approximately
60 cc/minute dry gas, or one std tank per month. Minimum temperature of 4ºC requires a laboratory ambient temperature
of 24ºC or below.
††† Minimum volume specified with 2 mg/mL Lysozyme in a 1536 well plate, Wyatt P/N P8803-1536, with 25 seconds acquisition.
Wyatt Technology is the world leader in providing the most advanced macromolecular and nanoparticle characterization tools. These instruments and software may be used to determine the absolute molecular weight and or size of macromolecules and nanoparticles and include: multi-angle light scattering, dynamic light scattering, high through-put dynamic light scattering, field flow fractionation, refractive index, zeta potential and viscometry detection.